Display cumulated views for phased-array inspections
When performing a phased-array inspection involving different sequences or shots for one or several mechanical scanning position, you can cumulate the views from all “Scan” or true “Bscan” in a single image.
To activate this cumulated view, just select the “data tools” option, available in the title bar of the true Bscan or Sscan image, and click on the “Cumulated view” icon:
For both of these options, the resulting image contains the complete data displayed in the image. “Merged” image means that, for each pixel of the image, the amplitude displayed corresponds to the maximum amplitude obtained in this region (if several shots correspond to this location), whereas “Mean Merged” image corresponds to the average of all shots for each pixel of the image.